AI may enter your QC lab with a software release

Waters has announced AI-powered peak integration for Empower. Five checks I would put in the qualification plan before a lab switches it on.

On 28 October 2026 Waters announced AI-Powered Peak Integration for Empower, its chromatography data system. Early access is planned for December 2026, with validated deployment targeted for mid-2027.1

No new instrument arrives. The capability comes as a subscription tier in software the lab already runs. Your lab still needs to show that it works for its intended use.

This article shows where the feature sits on the path from vial to result, then sets out the five checks. It rests on Waters’ announcement and on a draft regulatory text. At the time of writing, the feature itself is not yet available to test.

01 / The step

Where integration sits

A chromatography run produces a trace: detector signal against time. Integration turns that trace into numbers. The software decides where each peak starts and ends and where the baseline runs beneath it. The area between trace and baseline is what the result is calculated from.

The sequence below follows one set of samples from the vial to the reported result. It plays on its own. Pause it, drag the slider, or select any component marked with a plus to enlarge it.

Fig. 1 · From vial to resultIllustrative sequence
An illustration, not Empower. A generic ion-chromatography sequence for chloride in water, with a generic data system on the screen. Arrival times follow US EPA Method 300.0.6 Peak sizes, the calibration and the check limit are invented. Open the sequence on its own page
  1. Vial
  2. Loop
  3. Column
  4. Detector
  5. Trace
  6. Integration
  7. Calibration
  8. Result against the limit

Integration is the step the new feature takes over. The calibration line, the reported concentration and the verdict against the limit are all calculated from the areas it produces.

02 / The risk

Same trace, two results

The peaks in that sequence stand well apart. Routine data is often less tidy. The same chromatogram can yield a different peak area if the integration baseline changes.

Fig. 2 · Same trace, two baselinesSimulated data

Drop line to one baseline

One baseline runs under both peaks. A vertical line splits them.

Drop line to one baselineDetector signal in mAU against time from 4.1 to 6.1 minutes. A caffeine peak near 4.7 minutes is followed by a smaller neighbouring peak near 5.2 minutes on a slowly rising baseline. With this baseline the caffeine peak area is 11.8 mAU·min, reported as 3.93 mg/L.010203040504.55.05.56.0Signal (mAU)Time (min)CaffeineNearby compound
Caffeine reported3.93 mg/L
  • Peak area 11.8 mAU·min
  • 98% of the 4.00 mg/L in the vial
  • Inside a 90–110% limit

Valley-to-valley baseline

The baseline is lifted to the valley between the peaks.

Valley-to-valley baselineDetector signal in mAU against time from 4.1 to 6.1 minutes. A caffeine peak near 4.7 minutes is followed by a smaller neighbouring peak near 5.2 minutes on a slowly rising baseline. With this baseline the caffeine peak area is 10.7 mAU·min, reported as 3.57 mg/L.010203040504.55.05.56.0Signal (mAU)Time (min)Hatched: left out of both areasCaffeineNearby compound
Caffeine reported3.57 mg/L
  • Peak area 10.7 mAU·min
  • 89% of the 4.00 mg/L in the vial
  • Outside a 90–110% limit
Simulated trace. 4.00 mg/L caffeine and a tailing neighbour at resolution 1.5 on a drifting baseline, with a calibration of 3.0 mAU·min per mg/L. Move the pointer across either plot, or focus it and use the arrow keys, to read the signal.
Show the numbers as a table
Baseline rulePeak area, mAU·minReported, mg/LShare of the true 4.00 mg/LAgainst a 90–110% limit
Drop line to one baseline11.83.9398%Inside
Valley-to-valley baseline10.73.5789%Outside

Both panels show the same simulated trace: a 4.00 mg/L caffeine peak with a tailing neighbour on a drifting baseline. Drop a line between the peaks to a single baseline, and caffeine is reported at 3.93 mg/L. Lift the baseline to the valley between them, and it is reported at 3.57 mg/L. Against a 90–110% limit, the first result passes and the second fails.

Today the validated method fixes the rule. A model that takes over this step has to be shown to give equivalent areas, above all on traces like this one. So alongside routine runs, test the chromatograms your analysts already find difficult.

03 / The announcement

What the announcement says, and what it leaves open

Three statements in the release matter for a qualification plan.1

  • The models are “trained on each laboratory’s own data”.
  • They are version-controlled, and they keep their performance until the laboratory chooses to retrain and approve a new model.
  • The tasks named are removing noise peaks, combining high-molecular-weight peaks and editing baselines.

The performance figures come with different amounts of support.

Up to 95% fewer manual interventions
A pilot with one large pharmaceutical customer. 400 chromatograms were reviewed side by side: 90 interventions under current practice, 4 with the feature.
72% lower user-to-user variability and 50% faster review
Attributed to user testing. No sample size or method is given.
76% greater accuracy
Stated in the summary. No basis is given.

The release cites two Waters studies whose data is held on file. It names no regulation and does not say what the customer has to validate. That is normal for a launch announcement. It also means the figures cannot stand in for a lab’s own testing. If each model is trained on the laboratory’s own data, the model your lab would use has not been trained yet, let alone tested.

04 / The plan

Five checks for the qualification plan

The EU GMP Annex 22 consultation draft is a starting point.2 Here are five checks I’d put in the qualification plan.

  1. Intended use

    Define the integration task, the methods and sample types in scope, and the limits.

    Draft Annex 22 · 3.1The intended use is to be described in detail, the input data characterised and limitations identified. The description is approved before acceptance testing starts.

  2. Model version

    Confirm what stays fixed during use and who approves a retrained model.

    Draft Annex 22 · Scope and 10.2The draft covers static models that return the same output for the same input, placed under configuration control before use. A direct test: process the same chromatograms twice with one model version and compare the areas.

  3. Supplier evidence

    Identify and review the documentation relevant to your intended use.

    Draft Annex 22 · 2.2Documentation is to be available to the regulated user and reviewed by them, whether the model was developed in-house or supplied.

  4. Acceptance criteria

    If the model replaces a process, establish how that process performs today and set criteria at least as demanding.

    Draft Annex 22 · 4.3Acceptance criteria are to be “at least as high as the performance of the process it replaces”. That requires knowing how today’s integration performs before the comparison starts.

  5. Change control

    Detect unauthorised configuration changes, assess the need for retesting, and document the rationale for that decision.

    Draft Annex 22 · 10.1 and 10.2Changes to the model, the system or the process are evaluated to decide whether retesting is needed. A decision not to retest has to be justified.

The draft asks for more than these five: test data kept independent of training data (sections 5 and 6), explainability and confidence scores (8 and 9), monitoring in routine use (10.3 and 10.4) and records of human review (10.5).

05 / The draft

How firm is Annex 22?

Annex 22 is not in force. The consultation ran from 7 July to 7 October 2025.3 In November 2026 the published EU GMP guide still lists no Annex 22, and Annex 11 on computerised systems is still the 2011 revision.4 In June 2026 EMA held an expert workshop on whether adaptive and probabilistic models could be accommodated, so the scope may change.5

That does not make the draft less useful as a checklist. Intended use, acceptance criteria and change control are questions a validation team can answer for this feature today, under the procedures it already has for computerised systems.

Portrait of Miro Samija

Miro SamijaSenior Validation & CQV Lead · GMP Compliance & AI Governance

Fifteen years of GMP and CQV experience in pharma and biotech, from Pfizer Croatia to Merck’s Biotech Development Center in Switzerland. More about Miro

Sources

  1. Waters Corporation, news release announcing AI-Powered Peak Integration for Empower Software, 28 October 2026. prnewswire.com
  2. European Commission, EudraLex Volume 4, Annex 22: Artificial Intelligence, draft for stakeholders’ consultation, July 2025. health.ec.europa.eu (PDF)
  3. European Commission, Stakeholders’ Consultation on EudraLex Volume 4: Chapter 4, Annex 11 and New Annex 22, open 7 July to 7 October 2025. health.ec.europa.eu
  4. European Commission, EudraLex Volume 4: Good Manufacturing Practice guidelines, list of chapters and annexes, checked on 25 November 2026. health.ec.europa.eu
  5. European Medicines Agency, Multistakeholder workshop on expert contributions to artificial intelligence guidance development (Annex 22), 30 June and 1 July 2026; report published 2 October 2026. ema.europa.eu
  6. US Environmental Protection Agency, Method 300.0: Determination of Inorganic Anions by Ion Chromatography, Revision 2.1, 1993. epa.gov (PDF)

Waters and Empower are trademarks of Waters Corporation. This article is independent: it was not sponsored, reviewed or endorsed by Waters. Figures 1 and 2 are illustrations. The sequence uses invented peak sizes and limits, and the trace in Fig. 2 is simulated.

06 / The question

I co-authored the qualification plan for a programme covering more than 1,000 laboratory instruments. The plan determines what gets tested.

What would your lab test before enabling AI peak integration?